PROTEK 51-65 WATT MEDICAL SWITCHING POWER SUPPLIES
DESCRIPTION
The PM66 series of compact, open PCB constructed,
AC-DC switching power supplies are capable of delivering
51 to 65 watts of continuous power. They operate at 90 to
264VAC input voltage without the need of voltage
selection. They are ideally suited for use in medical
equipment not for life support. All models meet the safety
requirements of UL, CSA and IEC.
PM66 SERIES
( LV D )
RoHS
FE AT U R E S
Recognized or certified by UL,CSA and TÜV
Small size
100% burn-in
Wide input range 90 to 264VAC
Input surge current protection
Overvoltage protection
Overcurrent protection
Compliant with RoHS requirements
Safety Standard Approvals
INPUT SPECIFICATIONS
Input voltage
90 to 264VAC
Input frequency
47 to 63 Hz
Input current
1.60Armsfor 115VAC
1.00Armsfor 230VAC
Earth le aka ge cu rren t90 uA m ax. @ 115VAC, 6 0H z
150uA max. @ 230VAC, 50Hz
UL 2601-1, CSA C22.2 NO. 601.1
File No. E178020
TÜV EN60601-1
OUTPUT SPECIFICATIONS
Output voltage/currentSee rating chart
Total output power
65 watts maximum
Ripple and noise
1peak to peak maximum
Overvoltage protectionProvided on output; set at
112 - 132of its nominal
output voltage
Overcurrent protectionThe output protected to short
circuit conditions
Temperature coefficientAll outputs ±0.04/
maximum
Transient response
Maximum excursion of 4
or better on all models,
recovering to 1of final value
within 500us after a 25
step load change
EN V IR O NM ENTAL S P EC IF IC AT IO N S
Operating temperature: 0to +70
Storage temperature-40to +85
R e l a t i v e h u m i d i t y
5to 95non-condensing
Derating
Derate from 100at +50
linearly to 50at +70
2-11
GENERAL SPECIFICATIONS
Switching frequency42KHz ±5KHz
E f f i c i e n c y
75minimum on models with
Vo12V, 66% minimum on the others
Hold-up time
10 msec minimum at 110VAC
Line regulation
Inrush current
Withstand voltage
MTBF
±0.5maximum at full load
17 amps @ 115VAC or 40 amps
@230VAC, at 25cold start
4000VAC from input to output
1500VAC from input to ground
500VAC from output to ground
400,000 hours minimum at full load
at 25ambient , calculated per
MIL-HDBK-217F
EMC Performance (EN60601-1-2: 2001)
EN55011:
Class B conducted, Class B radiated
FCC:
Class B conducted, Class B radiated
VCCI:
Class B conducted, Class B radiated
EN61000-3-2:
Harmonic distortion, Class A
EN61000-3-3:
Line flicker
EN61000-4-2:
ESD, ±8KV air and ±6KV contact
EN61000-4-3:
EN61000-4-4:
Radiated immunity, 3V/m for
80~2500MHz
Fast transient/burst, ±2KV
EN61000-4-5:
Surge, ±1KV diff., ±2KV com.
EN61000-4-6:
Conducted immunity, 3Vrms
EN61000-4-8:
Magnetic field immunity, 3A/m
E N 6 1 0 0 0 - 4 - 11 :
Voltage dips, 30% reduction for 500ms,
60% reduction for 100ms and >95%
reduction for 10ms