DatasheetQ Logo
Electronic component search and free download site. Transistors,MosFET ,Diode,Integrated circuits

GHA500F-15-T3 데이터 시트보기 (PDF) - Unspecified

부품명
상세내역
제조사
GHA500F-15-T3
ETC
Unspecified ETC
GHA500F-15-T3 Datasheet PDF : 4 Pages
1 2 3 4
AC-DC Power Supplies Medical type
GHA500F
Ordering information
GH A 500 F
1
2
3
4
-OO -O
5
6
Recommended EMI/EMC Filter
R
EAC-10-472
Information the Home page is the latest.
Photo is an image of the appearance.
May differ from the actual product.
High voltage pulse noise type : EAP series
Low leakage current type : EAM series
*The EMI/EMC Filter is recommended to
connect with several devices.
Cautions
 ・Forced air cooling is required for the maximum output power. Please see instruction manual.
 ・Avoid applying stress to surface mount components.
 ・De-rating is required if the applied input voltage is 90-115VAC.
 ・The electrolytic capacitor has limited life span which is very much dependent on the actual operating conditions.
 ・Operating in the presense of chemical vapors or harsh environmnet can affect the power supply life expectnacy.
 ・Please make sure to read the instruction manual carefully before using this product.
It should be in the "Instruction Manual" not spec sheet.
MODEL
MAX OUTPUT WATTAGE[W]
Forced air at 50C
DC OUTPUT
Convection
at 40C
at 50C
conduction at 0C
cooling at 50C
GHA500F-12
500.8
12V 41.7A
12V 12.5A
12V 9.2A
12V 30.0A
12V 16.7A
GHA500F-15
501
15V 33.4A
15V 10.0A
15V 7.4A
15V 24.0A
15V 13.4A
GHA500F-24
504
24V 21.0A
24V 6.3A
24V 4.6A
24V 15.0A
24V 8.4A
1Series name
2Single output
3Output wattage
4Universal input
5Output voltage
6 Optional *6
T3 : mounting hole M3
J1 : VH(J.S.T.)connector type
R3 : with Subfeatures
(5VAUX,12VAUX,Remote,
Power good)
P : Pallarel Operation
Specification is changed at
option, refer to Instruction
manual.
GHA500F-48
504
48V 10.5A
48V 3.2A
48V 2.3A
48V 7.5A
48V 4.2A
SPECIFICATIONS
GHA
MODEL
GHA500F-12
GHA500F-15
GHA500F-24
GHA500F-48
VOLTAGE[V]
AC90 - 264 1f (output derating is required at AC90V -115V *3)
CURRENT[A]
ACIN 120V 5.4typ
ACIN 230V 2.9typ
FREQUENCY[Hz]
50 / 60 (47 - 63)
INPUT
EFFICIENCY[%]
ACIN 120V 88typ
ACIN 230V 90typ
90typ
92typ
90typ
92typ
90typ
92typ
POWER FACTOR ACIN 120V 0.95typ
(Io=100%) ACIN 230V 0.90typ
INRUSH CURRENT[A]
ACIN 120V
ACIN 230V
20typ (Io=100%) (At cold start) (Ta=25C)
40typ (Io=100%) (At cold start) (Ta=25C)
LEAKAGE CURRENT[mA] 0.125/0.250max (ACIN 120V/240V 60Hz,Io=100%, According to IEC60601-1)
VOLTAGE[V]
12
15
24
48
Forced air 41.7
33.4
21.0
10.5
CURRENT[A]
Convection 9.2
7.4
4.6
2.3
conduction cooling 16.7
13.4
8.4
4.2
LINE REGULATION[mV] *4 48max
60max
96max
192max
LOAD REGULATION[mV] *4 100max
RIPPLE[mVp-p]
*1
0 to +50C 240max
-20 - 0C 320max
120max
240max
320max
150max
240max
320max
240max
300max
400max
OUTPUT
RIPPLE NOISE[mVp-p]*1
0 to +50C 300max
-20 - 0C 360max
TEMPERATURE REGULATION[mV]
0 to +50C 120max
-20 to +50C 150max
300max
360max
120max
150max
300max
360max
240max
290max
480max
500max
480max
600max
DRIFT[mV]
*2 48max
60max
96max
192max
START-UP TIME[ms]
500typ (ACIN 120V, Io=100%)
HOLD-UP TIME[ms]
16typ (ACIN 120V, Io=100%)
OUTPUT VOLTAGE ADJUSTMENT RANGE[V] 10.80 to 13.20
13.50 to 16.50
21.60 to 26.40
43.20 to 52.80
OUTPUT VOLTAGE SETTING[V] 12.00 to 12.48
15.00 to 15.30
24.00 to 24.96
48.00 to 49.92
OVERCURRENT PROTECTION Works over 105% of rating and recovers automatically
PROTECTION
CIRCUIT AND
OTHERS
OVERVOLTAGE PROTECTION[V]
AUX1 (12V1A)
AUX2 (5V1A)
REMOTE ON/OFF
13.80 to 16.80
Optional
Optional
Optional
17.25 to 21.00
27.60 to 33.60
55.20 to 67.20
PowerGood
Optional
INPUT-OUTPUT-RC-AUX *7 AC4,000V 1minute, Cutoff current = 10mA, DC500V 50MW min (At Room Temperature)
ISOLATION
INPUT-FG
OUTPUT-RC-AUX-FG
AC2,000V 1minute, Cutoff current = 10mA, DC500V 50MW min (At Room Temperature)
*7 AC500V 1minute, Cutoff current = 25mA, DC500V 50MW min (At Room Temperature)
OUTPUT-RC-AUX
*7 AC500V 1minute, Cutoff current = 25mA, DC500V 50MW min (At Room Temperature)
OPERATING TEMP.,HUMID.AND ALTITUDE -20 to +80C, 20 - 90%RH (Non condensing)
ENVIRONMENT
STORAGE TEMP.,HUMID.AND ALTITUDE
VIBRATION
-30 to +80C, 20 - 90%RH (Non condensing)
10 - 55Hz, 19.6m/s2 (2G), 3minutes period, 60minutes each along X, Y and Z axis
IMPACT
196.1m/s2 (20G), 11ms, once each X, Y and Z axis
SAFETY AND AGENCY APPROVALS
UL60950-1, ANSI/AMII ES60601-1, C-UL, EN60950-1, EN60601-1
NOISE
CONDUCTED NOISE
Complies with FCC-B, VCCI-B, CISPR11-B, CISPR22-B, EN55011-B, EN55022-B
REGULATIONS HARMONIC ATTENUATOR
OTHERS
CASE SIZE/WEIGHT
COOLING METHOD
Complies with IEC61000-3-2 (class A) *5
76.2X35X127mm [3.0X1.4X5.0 inches] (WXHXD) / 420g max
Convection, Forced air (Require external fan), Conduction cooling
*1 This is the value that measured on measuring board with capacitor of 22mF at 150mm from
output terminal.
Measured by 20MHz oscilloscope or Ripple-Noise meter (Equivalent to KEISOKU-GIKEN: RM103).
*2 Drift is the change in DC output for an eight hour period after a half-hour warm-up at 25C, with
the input voltage held constant at the rated input/output.
*3 Derating is required.
*4 Please contact us about dynamic load and input response.
GHA-4
*5 Please contact us about another class.
*6 Specification is changed at option, refer to Instruction Manual.
*7 Applicable when AUX and remote control (optional) is added.
* To meet the specifications. Do not operate over-loaded condition.
* Sound noise may be generated by power supply in case of pulse load.
* Parallel operation is not possible.

Share Link: 